Configuration

Configuration of Qure.ai App with Medical System (modality)

Prerequisite

The following are the supported modalities.

  • X-Rays

  • CT scans

Setting up the Medical System (modality) Configuration

Before configuring the Qure.ai app, the user must have basic knowledge of modality networking to configure with Qure app successfully. Each modality is a computer that is linked to another computer through a network. The modality has its settings (local settings). The user must add the following settings in their modality:

  • AE Title: QUREAI

  • IP: Static IP of the system where the Qure.ai app is installed.

    Note: For more information on how to set up a static IP address in Windows, see How to set a static IP address in Windows.

  • Port: 5253

The modality must know the remote settings of the Qure.ai app with which it wants to communicate with.

Configuring Medical System (modality)

The main objective of this configuration is to initiate communication between the medical system and the Qure.ai app.

The following are the steps to configure the medical system (modality).

Attention:

The console will differ based on Vendor. Therefore, the steps to configure the nodes may also differ between modalities. So, consider the steps below as a reference example to configure any modality.

  1. Login to the modality console with an admin access privilege.

  2. On the modality console, select DICOM in the console configuration.

  3. On the Add/Remove: Archive, type the name/AE Title of the Qure.ai app. Then click Add.

  4. On the Configurations of the Archive, type the Alias name, IP-Address, AeTitle, and Port of the Qure.ai app as shown.

  5. Select Always send the images to this archive? checkbox. This option enables the modality to auto-push the study images to the Qure.ai app.

  6. After successfully configuring the modality console, click check connection to check the connection between the modality and Qure.ai app.

    Note: The option for checking the connection varies between vendors, for example, PING PACS workstation/RIS/ECHO Dicom/check connection.

  7. If the connection between them is Successful/Connected, the connection is complete.

  8. If the connection failed, repeat steps 4 through 6. If the issue still persists, seek the assistance of your modality vendor or engineer for support.